| MODEL |
DP3000 |
DP3001 |
| Attenuation Ratio: |
1/1,1/10,1/100 |
1/1,1/10,1/100 |
| Bandwidth: |
DC to 100MHz(-3dB) |
DC to 35MHz(-3dB) |
| BW limited: |
DC to 20MHz(-3dB) |
|
| Accuracy: |
3% |
3% |
| Rise time |
3.5ns |
10ns |
| 17.5ns(BWL) |
|
| Input impedance: |
1/1,1/10:10Mohm//5pF |
1/1,1/10:10Mohm//5pF |
| 1/100:11Mohm//5pF |
1/100:11Mohm//5pF |
| Diff. DC-input Resistance: |
>=20Mohm |
>=20Mohm |
| Max. Diff. Input Voltage: |
DC+AC 1/1 +/-10V |
DC+AC 1/1 +/-10V |
| 1/10 +/-100V |
1/10 +/-100V |
| 1/100 +/-1000V |
1/100 +/-1000V |
| Max. Common mode voltage |
1/1,1/10 +/-100V |
1/1,1/10 +/-100V |
| 1/100 +/-1000V |
1/100 +/-1000V |
| Output |
=<+/-10V |
=<+/-10V |
| Output impedance |
50ohM |
50ohm |
| CMRR |
1MHz>40dB |
1MHz>40dB |
|
 |

|
|
| [ENLARGE] |
|
| |
SPECIFICATIONS
Differential Voltage DC + pk AC: 6500V
Bandwidth ( 50次load -3dB ): 100MHz( x100 50MHz)
Common Mode Voltage DC + pk AC: 6500V
Common Mode Voltage RMS CAT II: 6500V
Common Mode Voltage RMS CAT III: 6500V
Attenuation ( Switchable ): X 100, X 200, X 500, X 1000
Input R ( Each input ): 27M次± 1%
Input C ( Each input ): 2.5 PF ± 2%
Maximum Operation Voltage( DC + pk AC )
汨± 650V at X 100
汨± 1300V at X 200
汨± 3250V at X 500
汨± 6500V at X 1000
Common Mode Rejection Ration
( C M R R )
60 Hz : > 20,000 : 1
100 Hz : > 2,000 : 1
1 MHz : > 600 : 1
Noise ( Into 50 O load ): 汨1 m Vrms
Input Impedance ( between input): 54 M次// 1.25 PF
Accuracy ( at 20~30⊥ 70%RH after 20 minutes ):汨±1.5 %
Maximum Output Voltage: 汨± 6.5 V
POWER SOURCE:
1. 9V battery
2. External 6V ~ 9V DC
FEATURES
- Do not use TDP100 above 6500V ( DC + peak AC ) between ground and the input or 6500V ( DC + peak AC ) between the input lead.
- The TDP100 converts the high voltage differencial input signal to a low voltage ground refereced signal for display on any Oscilloscope.
- The output BNC of TDP100 is calibrated to drive a high impedane ( 1 M次) load.
INSTRUCTION FOR USE
- Connect the output BNC of TDP100 to the input BNC of the Oscilloscope by the accessory BNC cable.
- Adjust the vertical offset of the Oscilloscope if necessary.
- Set the select proper range of the TDP100 and the V / DIV of the Oscilloscope according to the scale conversion chart.NOTE: If the voltage of the input signal exceeeds the linear range of the setting range. The signal output of the TDP100 would not accurately, the wave form display will be cut off.
- Scale conversion chart: The effective V / DIV is the attenuation factor of X 100 ~ X 1000 multiplied by the scale factor of the Oscilloscope. It will be twice when the 50O load was used. For example, with the range set at X 200, and the scope set to 0.5V / DIV, the effective V / DIV equals 200 X 0.5 or 100V, when the 50次load was used, it becomes 200V, the power consumption will increase too .
|
|
|