MODEL DP3000 DP3001
Attenuation Ratio: 1/1,1/10,1/100 1/1,1/10,1/100
Bandwidth: DC to 100MHz(-3dB) DC to 35MHz(-3dB)
BW limited: DC to 20MHz(-3dB)  
Accuracy: 3% 3%
Rise time 3.5ns 10ns
17.5ns(BWL)  
Input impedance: 1/1,1/10:10Mohm//5pF 1/1,1/10:10Mohm//5pF
1/100:11Mohm//5pF 1/100:11Mohm//5pF
Diff. DC-input Resistance: >=20Mohm >=20Mohm
Max. Diff. Input Voltage: DC+AC 1/1 +/-10V DC+AC 1/1 +/-10V
1/10 +/-100V 1/10 +/-100V
1/100 +/-1000V 1/100 +/-1000V
Max. Common mode voltage 1/1,1/10 +/-100V 1/1,1/10 +/-100V
1/100 +/-1000V 1/100 +/-1000V
Output =<+/-10V =<+/-10V
Output impedance 50ohM 50ohm
CMRR 1MHz>40dB 1MHz>40dB

 

4 RANGE ATTENUATOR
2 POWER SOURCE
6500V HIGH VOLTAGE INPUT.
<6.5V LOW BATT. INDICATE
<5.5V AUTO POWER OFF
SEPARATING DESIGN, CONVENIENT, DURABLE

MEET IEC1010, UL3111, TUV GS, CAT III 6500V

[ENLARGE]  
 

SPECIFICATIONS
Differential Voltage DC + pk AC: 6500V
Bandwidth ( 50次load -3dB ): 100MHz( x100 50MHz)
Common Mode Voltage DC + pk AC: 6500V
Common Mode Voltage RMS CAT II: 6500V
Common Mode Voltage RMS CAT III: 6500V
Attenuation ( Switchable ): X 100, X 200, X 500, X 1000
Input R ( Each input ): 27M次± 1%
Input C ( Each input ): 2.5 PF ± 2%
Maximum Operation Voltage( DC + pk AC )
汨± 650V at X 100
汨± 1300V at X 200
汨± 3250V at X 500
汨± 6500V at X 1000
Common Mode Rejection Ration
( C M R R )
60 Hz : > 20,000 : 1
100 Hz : > 2,000 : 1
1 MHz : > 600 : 1
Noise ( Into 50 O load ): 汨1 m Vrms
Input Impedance ( between input): 54 M次// 1.25 PF
Accuracy ( at 20~30⊥ 70%RH after 20 minutes ):汨±1.5 %
Maximum Output Voltage: 汨± 6.5 V
POWER SOURCE:
1. 9V battery
2. External 6V ~ 9V DC

FEATURES

  1. Do not use TDP100 above 6500V ( DC + peak AC ) between ground and the input or 6500V ( DC + peak AC ) between the input lead.
  2. The TDP100 converts the high voltage differencial input signal to a low voltage ground refereced signal for display on any Oscilloscope.
  3. The output BNC of TDP100 is calibrated to drive a high impedane ( 1 M次) load.
INSTRUCTION FOR USE
  1. Connect the output BNC of TDP100 to the input BNC of the Oscilloscope by the accessory BNC cable.
  2. Adjust the vertical offset of the Oscilloscope if necessary.
  3. Set the select proper range of the TDP100 and the V / DIV of the Oscilloscope according to the scale conversion chart.NOTE: If the voltage of the input signal exceeeds the linear range of the setting range. The signal output of the TDP100 would not accurately, the wave form display will be cut off.
  4. Scale conversion chart: The effective V / DIV is the attenuation factor of X 100 ~ X 1000 multiplied by the scale factor of the Oscilloscope. It will be twice when the 50O load was used. For example, with the range set at X 200, and the scope set to 0.5V / DIV, the effective V / DIV equals 200 X 0.5 or 100V, when the 50次load was used, it becomes 200V, the power consumption will increase too .