| PASSIVE PROBE (TEXAS-xx/I series) |

TEXAS20/I, TEXAS40/I, TEXAS60/I,TEXAS100/I |
Texas's
latest design of Passive Oscilloscope Probes feature a one piece (monolithic)
design for robust probe life
- Probe bandwidth ratings are specified as system bandwidth. System
bandwidth includes both probe and oscilloscope together as a system.
The result is a probe frequency response that essentially can be considered
transparent.
- Full set a accessories are included with each probe.
- All probes meet IEC1010-2-031,600Vpp, CATII
|
TEXAS xx/I Series
| Model#: |
Attenuator |
System
Bandwidth
(3dB)MHz: |
System
Rise-time
(ns) |
Compensation
Range (pf): |
Voltage
Max.
(Note2) |
Input Res.
(M ) |
Input Cap.
(Pf): |
GND.
Ref: |
| TEXAS20/I |
X1/X10 |
15/20 |
23.3/17.5 |
10-60 |
600 |
1/10 |
60/15 |
X |
| TEXAS40/I |
X1/X10 |
15/40 |
23.3/8.75 |
10-60 |
600 |
1/10 |
60/15 |
X |
| TEXAS60/I |
X1/X10 |
15/60 |
23.3/5.8 |
10-60 |
600 |
1/10 |
57/12 |
X |
| TEXAS100/I |
X1/X10 |
15/100 |
58/3.5 |
10-60 |
600 |
1/10 |
50/15 |
X |
|
 |
TEXAS150/I, TEXAS200/I,TEXAS250/I TEXAS
xx/I Series
| Model#: |
Attenuator |
System
Bandwidth
(3dB)MHz: |
System
Rise-time
(ns) |
Compensation
Range (pf): |
Voltage
Max.
(Note2) |
Input Res.
(M ) |
Input Cap.
(Pf): |
GND.
Ref: |
| TEXAS150/I |
X1/X10 |
10/150 |
35/2.33 |
5-30 |
600 |
1/10 |
82/12 |
X |
| TEXAS200/I |
X1/X10 |
10/200 |
35/1.75 |
5-30 |
600 |
1/10 |
53/10 |
X |
| TEXAS250/I |
X1/X10 |
10/250 |
35/1.4 |
5-30 |
600 |
1/10 |
57/10 |
X |
|
| Low Frequency Trimmer Adjustment: |
Low frequency response can be matched to the oscilloscope by adjusting
the compensation trimmer on the head of probe. |
-
Connect
the probe to the oscilloscope and to a 1KHz. Square wave source, the
rise time should not exceed 10 microseconds (most oscilloscopes provide
a probe compensation output). For X1/X10 probes, switch to the X10 position.
- Set the oscilloscope to display two to three cycles and two to six
vertical divisions.
- Carefully adjust the trimmer tool to obtain the flattest tops to
the square waves displayed on the oscilloscope, see illustrations
|

LF COMPENSATION |
| PASSIVE PROBE (TEXAS-xx/II series) |

TEXAS150/II,TEXAS200/II,TEXAS250/II SERIES |
TEXAS xx/I Series
| Model#: |
Attenuator |
System
Bandwidth
(3dB)MHz: |
System
Rise-time
(ns) |
Compensation
Range (pf): |
Voltage
Max.
(Note2) |
Input Res.
(M ) |
Input Cap.
(Pf): |
GND.
Ref: |
| TEXAS150/II |
X1/X10 |
10/150 |
35/2.33 |
5-30 |
600 |
1/10 |
60/12 |
X |
| TEXAS200/II |
X1/X10 |
10/200 |
35/1.75 |
5-30 |
600 |
1/10 |
60/12 |
X |
| TEXAS250/II |
X1/X10 |
10/250 |
35/1.4 |
5-30 |
600 |
1/10 |
60/12 |
X |
|
NOTE: 1.) Measurements
based on 1M ohm,20 pf. Oscilloscope.
2.) Maximum probe input voltage is limited to the lesser of 600Vpp
3.) Model TEXAS150/I, TEXAS200/I, TEXAS250/I,TEXAS150/II,TEXAS200/II,TEXAS250/II
probes have two adjust resistance of BNC connector. If your probe must be
to adjust, please take it to careerman for assistance.
4.) Model TEXAS150/I, TEXAS200/I, TEXAS250/I ,TEXAS150/II,TEXAS200/II,TEXAS250/II,measurement
based on 1Mohm, 18pf oscilloscope |
| High Frequency Trimmer Adjustment: (Model
TEXAS150/I, TEXAS200/I and TEXAS250/I, TEXAS150/II, TEXAS200/II, TEXAS250/II)
The high frequency compensation has been pre-adjusted at the factory,
however if adjustment is required, use the following procedure. |
-
The
high frequency compensation box cover located near the connector, Use
the BNC adapter, connected the probe to a square wave generator operating
between 10 to 100KHz terminated into 50ohms. The square wave generator
rise time should be approximately.125ns. Adjust each control until the
leading edge of the waveform is as flat, square and horizontal as possible
- (see high frequency illustrations). The best
take it to careerman for assistance.
- Readjust the low frequency compensation if necessary.
|

HIGH FREQUENCY COMPENSATION |